• Lanaform® cannot be held liable in the event of accidental damage resulting from any use contrary to the instructions given in these instructions for use.

MAINTENANCE AND STORAGE

If the water contains too much calcium and magnesium, it may deposit a "white powder" in the device. If this foam is deposited on the transducer (ceramic ring producing ultrasounds,
diagram 1, point 3), this will prevent the device from working properly.
Maintenance:
We recommend:
1. Using cooled, boiled water for each use;
2. Cleaning the reservoir every 2 days and cleaning the transducer every week;
3. Changing the water in the reservoir every 2 days;
4. Cleaning the device and making sure that all its parts are completely dry when it is not in use.
To clean the transducer:
1. Switch the device off and unplug it;
2. Put 2 to 5 drops of vinegar on its surface and leave them there for 2 to 5 minutes;
3. Brush the foam off the surface with a soft brush (cotton bud) (21). Do not use a hard tool to scrape the surface;
4. Rinse the reservoir of the base in clean water;
5. Do not use soap, solvent or spray cleaner to clean the transducer.
To clean the reservoir:
1. Switch the device off and unplug it;
2. Clean the reservoir with water using a soft cloth and a little highly diluted vinegar if there are foam or limescale deposits on the inside. Do not use any corrosive detergents.
Rub the ceramic disc gently with a cotton bud, taking care not to scratch it; (22)
3. Rinse the reservoir in clean water.
Storage:
After cleaning and drying the device, store it in a dry, cool place out of the reach of children.

TROUBLESHOOTING

In the event of a malfunction, under normal conditions of use, please refer to the following chart:
Breakdowns
No mist
Mist has an unusual smell
Low intensity
Possible causes
The water level in the reservoir is too low.
The temperature of the unit is too low.
The water has been standing in the reservoir too long.
Too much/too little water in the reservoir.
Sediment on the transducer
The water is too cold.
The water is not clean.
Solutions
Add water to the empty reservoir.
Place the unit in a room at ambient temperature for half an hour
prior to use.
Clean the reservoir and pour fresh water into it.
Empty water out of the reservoir or add water to it.
Clean the transducer.
Use water that is at room temperature.
Clean the reservoir and pour fresh water into it.