A HEAT SHIELD/SHOWER PROTECTOR (6) (P/N 730, sold
separately) should be used as required with the inhalation opening facing
downward.
An ADAPTER (2) (P/N 707, 708, sold separately) is only required for
connection to a cannula (1) without an ISO 15 connector (taper).
SUPPLEMENTAL OXYGEN can connect via the Ø5.5 mm nipple on
the valve housing. The WHITE O2 NIPPLE CAP (7) (P/N 710/710R
only) shall be fitted to the Speaking Valve and removed only when
oxygen is being supplied.
The SPEAKING VALVE should be replaced with a new unit after
29 days of use, or more frequently if necessary. The valve must not
be used while the patient is sleeping or otherwise when the patient's
consciousness is impaired. Use a SPIRO HUMIDIFIER (P/N 702, sold
separately) in such cases.
The Speaking Valve must be removed before inhalation of medicine.
RESISTANCE TO FLOW for SPIRO Speaking Valve is < 2 cm H
O*,
2
and with the heat shield/shower protection on < 3 cm H
O*.
2
All products should be stored in a dark and dry environment and at
normal indoor temperature.
For the STORAGE PERIOD OF SPIRO PRODUCTS: See year and
month next to the 'HOURGLASS' symbol on the product label. The
products shall be disposed of in accordance with the applicable national
regulations.
ALWAYS CONSULT YOUR DOCTOR, SUPPLIER OR
MANUFACTURER IF YOU HAVE ANY QUESTIONS.
* at 30 litres per minute