Detailed Explanations On The Chapter: "Working With The Laboratory Hot Plates; Timer Functions - SI Analytics SLK 1 Manual De Instrucciones

Hornillo de laboratorio con superficie de vitrocerámica
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Detailed explanations on the Chapter: "Working with the laboratory
hot plates".
"On / Off" sensor key, for switching the laboratory hot plate on and off (On / Off switch).
- To switch on the laboratory hot plate, press this key for at least 2 seconds. An acoustic
signal will indicate successful operation.
- If no other sensor is touched within 10 seconds of switching on the laboratory hot plate,
the laboratory hot plate will automatically switch off.
- To switch off, press this key for at least 2 seconds. An acoustic signal will indicate
successful operation.
- When switching off the laboratory hot plate, all settings with the exception of those of
the "Egg timer" and an activated key lock will for safety reasons be switched down to
/
Sensor keys for increasing <
operating time.
- An acoustic signal confirms every successfully completed operation.
- If both keys are touched at the same time, the system will switch to << 0 >>.
- << H >> in the display indicates the presence of residual heat (see Chapter on Putting into
operation).
! Danger of burning !
-
Display of the heat intensity settings 0 ... 9.
- setting << 1 >> = lowest heating output
- setting << 9 >> = highest heating output
Lock keys:
By pressing the key <Lock keys>, the control system can be put into a safe operating mode; in this mode
no changes can be made to the heating levels.
In locked mode, the laboratory hot plate can only be switched off.
To lock the keys, press <Lock sensor key> for at least 2 seconds.
Locking is confirmed by an acoustic signal.
To indicate that the key is locked, a LED lamp will light up above <Lock sensor key>.
If a sensor key is pressed in this mode, there will be no reaction.
The key lock will remain activated after the laboratory hot plate has been switched off.
Unlock keys:
To unlock the control system again, press <Lock sensor key> for at least 2 seconds. Release is confirmed
by an acoustic signal.
The LED lamp above <Lock sensor key> goes out again.
The laboratory hot plate can now be used again without any restriction.

Timer functions

The laboratory hot plate contains a timer with the two following functions:
- A timer function for the heating zone. The heating zone automatically switches off after the time set in the
timer has expired. The timer for the heating zone can only be set, when a heat intensity between 1 and 9
has been selected.
- A freely available "egg timer" function. The "egg timer" function operates regardless of whether the
heating zone has been switched on or off. An acoustic signal is emitted when the set time has expired.
The time set on the "egg timer" does not in any way affect the operation of the heating zone.
In order to use either of the timer functions, the timer must first be activated by pressing the <Timer sensor
key>.
When the <Timer sensor key> is pressed for the first time, the "egg timer function" can be programmed.
To indicate this mode, only the LED lamp to the right of the timer time display will light up.
When the <Timer sensor key> is pressed for a second time and the heating zone is switched on, the
heating zone can be programmed. This mode is indicated by an additional flashing LED lamp to the left of
the timer time display.
19