Table 2 Guidance and Manufacturer declaration - electromagnetic immunity -
for all ME Equipment and ME System
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
IEC 60601 test level
±6 kV contact
±8 kV air
Compliance
Electromagnetic
level
environment - guidance
±6 kV contact
Floors should be wood,
±8 kV air
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.