Table 4 Guidance and Manufacturer's declaration - electromagnetic Immunity -
for ME Equipment and ME Systems that are not Life-Supporting
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
IMMUNITY
IEC 60601 TEST
test
LEVEL
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
98
Compliance
Electromagnetic immunity
level
Not applicable
Portable and mobile RF communications
Equipment should be used no closer to
any part of the device, including cables,
3 V/m
than the recommended separation distance
calculated from the equation applicable to
the frequency of
the transmitter.
Recommended separation distance