Mechanical Cleaning/Disinfecting - Braun Aesculap Acculan 3Ti Instrucciones De Manejo

Ocultar thumbs Ver también para Aesculap Acculan 3Ti:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 76
Aesculap
®
Acculan 3Ti
Acculan 3Ti small drill GA671
5.8

Mechanical cleaning/disinfecting

Note
The cleaning and disinfection device must be of tested
and approved effectiveness (e.g. FDA approval or CE
mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for process-
ing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Prerinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfect-
ing
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water
quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechani-
cal cleaning/disinfecting.
20
T
t
[°C/°F]
[min]
<25/77
3
55/131
10
>10/50
1
90/194
5
-
-
Water
Chemical/Note
quality
D–W
-
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
FD-W
-
FD-W
-
-
According to the program for cleaning
and disinfection device