Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

SP-04-2013 OK (830P.) EN-DE FR-ES-PL-RUS_830P. 13-04-13 15:11 Strona 8

6 - TROUBLESHOOTING

If during your work the functioning of the pipette controller is incorrect, check the cause
and remove the fault.
Problem
The pipette falls out (the
holding force of the pipette
is to small), or tilts to the
side too much.
The pump is working, but
the pipette controller aid
does not draw liquid or
draws liquid very slowly.
Liquid leaks from the
pipette (the aspiration
and the dispense buttons
are not pressed)
If the above actions do not help, the device shall be sent to HTL service.
Before that the product should be cleaned and decontaminated. Fill in the questionnaire
available at www.htl.com.pl with precise specification of solutions used and the type
of laboratory, in which the device has been used, and send it together with the product.
5
Possible cause
The pipette holder is dirty
or wet (fig. 1G).
The pipette holder is dam-
aged.
The filter is dirty (fig. 1H).
The pipette holder
and/or the connector gas-
ket is damaged (fig. 1J).
The pipette is damaged.
The pipette is inserted
incorrectly.
The pipette holder, the filter,
or the connector gasket
is installed incorrectly.
The pipette holder and/or
the connector gasket is
damaged (fig.1G, fig. 1J).
Action
Take out the pipette holder, clean,
wash and dry.
Replace the holder with a new
one.
Take out the pipette holder, take
out the filter, if it is dirty, replace it
with a new one.
Replace the mechanically dam-
aged elements with new ones.
Check the pipette for damage
- cracks, dents - if present, repla -
ce the pipette with a new one.
Check whether the pipette has
been correctly inserted in the
holder.
Check whether all parts are pre -
sent and correctly installed.
Replace the mechanically dam-
aged elements with new ones.

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Swiftpet pro

Tabla de contenido