The selected measured value is displayed for information (Fig. 122).
Fig. 122
«Measuring speed» or drift will determine the span of time after which the measured value will be accepted
following a titration step (Fig. 123).
Fig. 123
Drift-controlled acceptance of the measured value in terms of mV/min is set by selecting «normal», «fast» or
«user-defined» (Fig. 124).
The drift values at predefined in terms of in mV/min for normal and fast drift:
Normal Drift
Fast Drift
Small drift value
Large drift value
The following parameter selection can be made for user-defined drift setting:
Minimum holding time [s] 01 - 99
Maximum holding time [s] 01 - 99
Measuring time [s]
Drift [mv/min]
20 mV/min
50 mV/min
= slow and precise
= fast and "less precise"
01 - 99
01 - 99
183