238
Command
Description
aaAA
automatic allocation of device address
aaMC1...XX
choosing a method
"filling burette"
aaBF
aaBV
output of dosed volume in ml
aaDA
dose volume without filling, with adding the volume
aaDB
dose volume without filling, reset of the volume
aaDO
dose volume with filling, without adding the volume
aaGDM
dosing speed in ml/min
aaGF
filling time in seconds (min is 20, default 30)
"exit" function.back to main menu
aaEX
µa "dead stop" measurement function
aafd
aafp
pH measurement function
aaft
temperature measurement function
aafv
mV measurement function
dosing speed in ml/min (0.01 – 100 ml/min)
aagdm
filling time in sec (adjustable 20 – 999 seconds)
aaGF
aaGS
output serial no. Of device
aaLC
output of the CAL parameters
aaLD
output of the measurement data
aaLR
output report (short report)
aaM
output of the preset measurement value (pH/mV/µA)
aaRH
request of identification
aaRC
send last command
aaRS
report status
possible answers are:
"STATUS:READY" for ready
"STATUS:dosing" dosing
"STATUS:filling" filling
"ERROR:busy" if no interchangeable unit has been attached
aaSM
start selected method
aaSEEPROM
EEPROM reset to factory defaults
aaSR
stop the actual function
aaSS
titration start with the transfer of the pH end value
aaVE
Version number of the software
Reply
aaY
aaY
aaY
aa0.200
aaY
aaY
aaY
aaY
aaY
aay
aay
aay
aay
aay
aaY
aaY
aaGS08154711
aaY
aaY
aaM7.000
aaIdent: TL 7000
aa"last command"
aaStatus:"text
aaY
aaY
aaY
aaY
aaVersion