244
9.5
Autosampler
Connection of sampler changer TW alpha plus
9.5.1
The sample changer TW alpha plus is connected to the RS232-2 (RS2) of the titrator with cable TZ 3087.
The settings of the RS232-2 interface must be changed to 4800, No, 7, 2 (Fig. 281).
Fig. 281
The settings of the RS232-1 (4800, No. 8, 1) remain unaffected.
Connection of sample changer TW 7400
9.5.2
The sample changer TW 7400 plus is connected to RS232-2 (RS2) of the titrator by cable TZ 3987.
The settings of the RS232-2 interface do not have to be changed. They can remain at 4800, No. 8.1.
9.6
Using software TitriSoft
General
9.6.1
The titrator is connected to the PC via the RS232 or USB-1-B interface. Cables TZ 3097 and TZ 3091 can be
used via RS232-1 for the connection.
TitriSoft 3.15 or higher
9.6.2
When using the new software 3.15 or higher TitriSoft, the factory settings of the RS232-1 can be maintained.
Reading and writing the intelligent exchange units and ID electrodes is possible with TitriSoft 3.15.
For more information, please refer to the operating manual of TitriSoft.