Specifications
Specifications are subject to change without notice.
Measurement
Measurement method
Low measurement range
Measuring range
High measurement range
Display units
Reproducibility
Measurement accuracy
Limit of detection
Response time
Calibration
Sensor software
Software version
Equipment properties
Measurement sensor
Weight
Flow cell
Measurement sensor
Pressure range
Flow cell
Measurement sensor
(Ø × length)
Dimensions
Flow cell (L × W × D)
UV fluorescent measurement process for polycyclic aromatic hydrocarbons (PAH)
Excitation wavelength: 254 nm Emission (measurement) wavelength: 360 nm
0 to 50 ppb and 0 to 500 ppb in relation to PAH calibration standard, c orresponding to
0.1 to 1.5 ppm and 0.1 to 15 ppm of oil calibration standard
0 to 500 ppb and 0 to 5000 ppb in relation to PAH calibration standard, corresponding to
0.1 to 15 ppm and 0.1 to 150 ppm of oil calibration standard
ppb, ppm, µg/L, mg/L
2.5% of measurement value at a constant temperature
5% of measurement value ±2% from measurement range limit at a constant temperature
1.2 ppb (PAH)
10 s (T90)
Factory calibration with UV fluorescence calibration standard, c ustom adaptation possible
From 1.14
Stainless steel version 2.8 kg, titanium version 1.8 kg
Approximately 0.6 kg, including installation board approximately 2.0 kg
Max. 30 bar
Max. 1 bar
68 mm × 306 mm (2.68 in. × 12.05 in.) (without plugs or suspension pin)
68 mm × 399 mm (2.68 in. × 15.71 in.) (including suspension pin)
68 mm × 413 mm (2.68 in. × 16.26 in.) (with additional cleaning option)
98 mm × 98 mm × 150 mm (3.86 in. × 3.86 in. × 5.91 in.) (without fittings),
installation board: 600 × 300 × 10 mm (23.62 in. × 11.81 in. × 0.39 in.)
English