Function Test - HACH LANGE FP 360 sc Manual Básico Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 67

Function test

C A U T I O N
Before power is applied, refer to the controller operation
instructions.
After the components are removed from the package, do a function test.
1.
Connect the connector cable to the sensor (8-pin polarized
connector) and an appropriate sc controller (5-pin polarized
connector) (refer to
Connect sensor cable on page
2.
Apply power to the sc controller. The display is activated and the
sensor goes to measurement mode.
The sensor ticks quietly and regularly.
3.
Cover the sensor measurement window with a sheet of white paper
(do not use recycled paper).
4.
Vary the distance between the measurement window and the
paper. The measurement value on the display will change
accordingly.
Note: In air, the measurement value displayed is not exactly zero due to
reflections on the window surface (refer to
page
14).
Installation
D A N G E R
Personal injury hazard. Only qualified personnel should conduct
the tasks described in this section of the manual.
N O T I C E
If the sensor is not fully inserted, sun protection is recommended
in high ambient temperatures and intense solar radiation to
protect against thermal and UV effects.
English 8
8).
Check the zero point on
Connect sensor cable
C A U T I O N
Always put the cables and hoses in a position that does not bend
or cause a trip hazard.
C A U T I O N
Before power is applied, refer to the controller operation
instructions.
1.
Connect the polarized socket on the connector cable to the sensor
plug (8-pin plug) (refer to
2.
Push the safety sleeve on the plug (step 2).
3.
Hand-tighten the safety sleeve in position (step 3).
Figure 3 Connect the sensor cable to the sensor
Figure 3
step 1).

Publicidad

Tabla de contenido
loading

Tabla de contenido