BATTERY CHARGING

On 7305P series, the units are equipped with a factory-installed rechargeable battery. The unit will have a light for low battery and charge indication. On 7305D series, the
units are not factory equipped with a rechargeable battery; check with your equipment provider to determine if your unit has been upgraded with a rechargeable battery.
When you connect the unit to an AC or DC power source using the AC adapter or DC power cord, the green external power light will illuminate. The yellow charge indicator will
be illuminated while the battery is charging. Verify that this illuminates when charging begins. As the charge nears a full charge, the yellow LED may flash on and off for
several minutes. This is normal.
NOTE– Charge battery for a minimum of 17 hours before first use.
NOTE– Fully recharge battery after each use. The unit will continue to float charge the battery after the charge indicator turns off, so leave the unit connected to AC when not
in use.
CAUTION– Discharging the battery completely will shorten the life of the battery. Do not operate the unit more than a few minutes if the low battery indicator light is lit.
Recharge as soon as possible.
NOTE– Unit run time will decrease as the battery ages.
NOTE– Unit run time will also be reduced by letting the battery sit at a discharged state for extended periods.
STORAGE NOTE– Battery should be charged for a minimum of 17 hours prior to storage, and at least once every 6 months. Important - If battery recharge is delayed beyond
6 months, battery may be able to provide full run time after completion of 3 full charge and discharge cycles.
NOTE– A fully charged battery will provide approximately 60 minutes of continuous operation at a zero vacuum level (free flow). Operation time will decrease with higher
vacuum levels.
NOTE– When charging the battery, use an external power source and verify that the charge light illuminates when the unit is in the "Off' position. If the battery does not
charge, please be sure the model you are using has a battery installed prior to contacting your authorized DeVilbiss Healthcare provider.
NOTE– The internal rechargeable battery is sealed lead-acid. Contact your local authorities for instruction on proper disposal.
NOTE– Do not connect the AC adapter to an outlet controlled by a switch to ensure power is supplied to unit at all times.
NOTE– Do not connect the DC power cord to an outlet that is not constantly energized.

CLEANING INSTRUCTIONS

Preparation
1.
Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the DC input receptacle on the unit.
2.
Disconnect tubing and remove container from holder.
3.
Carefully remove lid and empty contents.
WARNING
To prevent possible risk of infection from contaminated cleaning/disinfection solutions, always prepare fresh solution for each cleaning
cycle and discard solution after each use.
Disposable Collection Container with Internal Filter Cartridge
The disposable collection container and lid are meant for single-patient use.
1.
Remove filter cartridge and 4-1/2" (11.43 cm) tubing and set aside.
NOTE– Filter MUST NOT get wet. The filter material cannot be removed from the elbow (figure A).
WARNING
Do not remove float ball from lid. If removed, float ball may pose a choking hazard (figure B).
2.
Wash container, lid and splash guard in warm water/dishwashing solution. Rinse with clean, warm water.
3.
Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes. Rinse with clean,
warm water and air dry.
NOTE– The disassembled container may also be washed in a dishwasher, top shelf only, using a cycle with a water temperature between
131ºF-149ºF/55ºC-65ºC.
Reusable Collection Container with External Bacteria Filter
1.
Remove bacteria filter, 4-1/2" (11.43 cm) tubing, and connection elbow and set aside. Remove o-ring and overflow valve from lid.
2.
Wash jar, lid, o-ring, and overflow valve in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3.
For single patient use: After washing, disinfect using one of the following methods:
a.
Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes. Rinse with clean, warm water and air dry
in a clean environment.
b.
Wash with rubbing alcohol and air dry in a clean environment.
c.
Wash with a commercial (bacterial-germicidal) disinfectant. Follow disinfectant manufacturer's recommended dilution rates and instructions carefully.
For multi-patient use: After washing:
4.
a.
After parts are completely dry, place jar and lid in autoclave with the open end down. Ensure parts are not touching. Run one sterilization steam cycle at 250°F
(121°C) for 15 minutes. NOTE– Jar is guaranteed up to 30 cycles of autoclave sterilization at the indicated conditions.
b.
Dispose of and replace filter, tubing and elbow between patients.