5. To adjust the voltage or current scaling for other than zero to full scale, Press and hold the MODE key while
pressing the FLOW key. This display will show "LO" for 2 seconds and then a flow rate. Use the UP/DOWN ( ,
) arrow keys to set the minimum control level. Press the FLOW key again. The display will show "HI" for 2
seconds and then a flow rate. Use the UP/DOWN ( ,
other key to exit. The same scaling will be used for both input and output levels.
NOTE:
Tubing
size
selected before adjusting the
voltage or current scaling.
PRIME
CW/CCW
should
be
RED/YELLOW
START/STOP
YELLOW
INPUT
0-20mA; 4-20mA
ORANGE
INPUT
0-10V
BROWN
OUTPUT
0-20mA; 4-20mA
OUTPUT
0-10V
) arrow keys to set the maximum control level. Press any
77420-00
14
NOTE:
BLUE
6
GREEN
4
10
8
12
WHITE
3
11
5
13
BLACK
1
15
7
16
VIOLET
9
17
RED
2
18
4
Colors are those of Remote Cable,
Catalog number 77300-32.
GREY
TACH OUTPUT
TAN
PINK
MOTOR RUNNING
N.O. CONTACT
RED/GREEN
MOTOR RUNNING
N.C. CONTACT
RED/BLACK
N.C.
N.C.
N.C.