Operation and function

Electronic thermostat
The appliance has an electronic thermostat
that controls the temperature in the vaccine
compartment. The thermostat is set by the
factory
1. The green lamp is on when the appliance
is connected to the power supply.
2. Thermometer.
Starting procedure
Connect the mains lead to the power
supply. The green lamp must light (fig. 4) to
indicate that the appliance is operating. The
electronic thermostat makes a self test, 20
seconds before the compressor starts.
Temperature control
It is recommended to check the inside
temperature with an accurate thermometer
once a day. The inside temperature must
be checked regularly according to WHO's
standards and specifications.

Cool down of the appliance

Before the appliance is loaded with
vaccines the ice-lining in the 4 sides must
be frozen. To ensure that the ice-lining in
the 4 sides is frozen do the following:
1. Place a thermometer in the top basket.
2. Plug in the appliance, and let it run for
24 h.
3. Check the temperature in the top and
bottom basket (must be between +2° and
+8°C).
The temperature in the vaccine
compartment must always be monitored
on the thermometer and be within the
range +2° to +8°C. Due to tolerance of the
thermostat you should always control the
temperature during cool down.