Maintenance and Cleaning of Infrared Touch Probe IRP40.02
Risk of damage to the equipment!
•
Clean and dry the probe well before opening!
•
Do NOT clean with compressed air or high pressure water!
•
Do NOT use any sharp tools (these could damage the inner seal)!
1.
Unscrew the stylus.
Unter dem Service-Deckel kann sich Schmutz ansammeln.
2.
Remove the service cover using the mounting tool (5840).
3.
Clean probe and parts under running water.
4.
Reassemble all parts.
5.
Align the stylus to spindle centre.
6.
Calibrate the probe!
O-Ring 23x1 Viton (6161)
Fig. 13
Maintenance and cleaning of infrared touch probe IRP40.02
NOTICE
INFORMATION
Mounting tool (5840)
Service-cover (5773)
Inner seal