RECOMMENDED SEPARATION DISTANCES

for portable an mobile RF communication equipment and the EUT.
The EUT is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or user of the EUT can help prevent electro-
magnetic interference by maintaining a minimum distance between portable and
mobile RF communication equipment (transmitters) and the EUT as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter [W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres [m] can be determined using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts [W] according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidances may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
184
Separation distance according to frequency of transmitter in [m]
150 kHz to 80 MHz
d = 3.5/3 * SQRT (P)
0.1
0.4
1.2
3.7
11.7
80 MHz to 800 MHz
d = 3.5/3 * SQRT (P)
0.1
0.4
1.2
3.7
11.7
Table 6 of DIN EN 60601-1-2:2007 (IEC 60601-1-2:2007)
800 MHz to 2.5 GHz
d = 7/3 * SQRT (P)
0.2
0.7
2.3
7.4
23.3