5. Calibration / Zero point calibration

MODEL
SCALES
RSD200
Refractive index
RSD500
Refractive index 1,3330-1,5100
Oe Germany
Oe (Swiss)
RSD205
Mass S/W
Vol AP
KMW
The zero position on the refractometer calibration must be checked before each use and afterwards.
A zero point calibration should be performed before any measurement where the highest possible accuracy is
required or if the environmental temperature has great variations. It is recommended to use deionized water.
(The temperature for the calibration liquid should be ideally near 20°C / 68°F)
-Inspect the prism to ensure that is clean and dry.
-Place a few drops of calibration liquid on the prism
-Press
(calibration button) to start
-The refractometer will show the calibration process on the LCD (Image 1)
-If the procedure is successful a "PASS" message will appear on the screen (Image 2)
-After the calibration, the LCD will return to temperature mode. Calibration result will be saved and it will be
the new zero point after the device is turned off and on again.
RANG
Brix
0-45%
1.3330-1.4098
Brix
0-85%
3-150 / 0-150
0-35% / 0-22%
0-25
Note:
-
After calibration, please remember to
clean the liquid from the prism surface and
keep the prism surface dry and clean to
avoid erosion.
Image 1
DIVISIONS ACCURACY
0.1
±0.1
0.0001
±0.0003
0.1
±0.5
0.0001
±0.0005
1
±1
1
±1
0,1
±0,2
0,1
±0,2
0,1
±0,2
5
CALIBRATION
Deionized water
Deionized water
Deionized water
Image 2