Caution The illuminator uses a highly concentrated light source (luminous power per area) and
this high energy density is retained through connected lightguides and instruments. The output of a
connected instrument left in close proximity or contact with tissue or flammable materials may present a
risk of injury or fire depending upon application. Qualified personnel must determine a safe working
distance and intensity setting for each application. The output should never be left on unattended.
Caution Use of this illuminator with accessories or attachments requires that the end-user follow
all accessory or attachment instructions which could affect illuminator setup, usage and/ or settings.
Caution For grounding reliability, use only hospital grade marked power cord and receptacle.
3. SPECIFICATIONS
Item
Lamp type
Power
Color temperature
Lamp life
Lamp replacement
Light guide adapter
Brightness control
Input voltage
Power consumption
Regulatory Approvals
Equipment Class
Mode of Operation
Water Resistant
Operating Environment
Temperature
Relative Humidity
Air Pressure
Storage Environment
Temperature
Relative Humidity
Air Pressure
Dimensions
Weight
LIT034
®
CUDA
SURGICAL ELITE 180
(English)
Specification
Cermax type 175 watt Xenon
175 Watt
5600 K
650 hours (typical)
Cartridge replacement from the side of the unit
Turret type to fit your choice of four: Storz, ACMI, Pentax, Wolf,
Olympus
Mechanical iris control from the front panel
100-120V AC, 50/60 Hz
220-240V AC,50 Hz
250 watt max
UL60601, EN 60601-1, EN 60601-1-2, CAN/CSA, C22.2,
No 601.1-M90
BF-type
Continuous operation
Not Protected Equipment, IPX0
+10° to +40° C (50° to 104° F)
30 to 85%
700 to 1060 hPa
-20° to +60° C (-4° to 140° F)
0 to 95%
700 to 1060 hPa
355(14)W x 130 (5)H x 250 (10)D mm
9.5 lbs./4.3 kg
Rev. P
Date of Revision: 02/09/17
Page 5 of 72