www.comfortseries.respironics.com
forehead Pad replacement
If you need to replace the forehead pad, contact your home
care provider to reorder. Follow the instructions below.
step 1
Remove the pad. Slide the pad outward toward
one edge of the forehead support, and
lift it upward so the connection track on that side
is released from the plastic connection slot. Next,
pull the pad in the opposite direction to release
the connection track on the other side from its
connection slot.
step 2
Install the new pad. Slide one of the connection tracks on the new pad into its connection slot
on the support. Pull the pad in the opposite direction (toward the center) and slide the other
connection track into its connection slot to lock the pad into place.
note: Replace the cushion and forehead pad if damaged, deformed, or hardened.
Disposal of Parts
Dispose of any damaged or broken parts in accordance with local regulations.
ComfortSelect Reorder Numbers
1033678 Premium Headgear w/ EZPeel™ Tabs
1002801 Deluxe Headgear
1016071 ComfortSelect Nasal Mask w/ hgr, S*
1016070 ComfortSelect Nasal Mask w/ hgr, M*
1016072 ComfortSelect Nasal Mask w/ hgr, S Wide*
1008463 ComfortSelect Nasal Mask w/ hgr, S**
1008462 ComfortSelect Nasal Mask w/ hgr, M**
1008464 ComfortSelect Nasal Mask w/ hgr, S Wide**
1007936 ComfortSelect Cushion, S
1007935 ComfortSelect Cushion, M
1007937 ComfortSelect Cushion, S Wide
* For ComfortSelect mask without exhalation elbow.
** For ComfortSelect mask with exhalation elbow.
Visit www.respironics.com for a complete list of products and reorder numbers, or contact Customer Service
at +1-724-387-4000 or +49 (0) 8152-93060 to obtain a product catalog (#1027825).
Deadspace
Pressure Drop (cm H
O/hPa)
2
S 119 cc
50 LPM
M 129 cc
All Cushion Sizes :
SW 128 cc
S-M-SW
0.1
© 2006 Respironics, Inc. and its affiliates. All rights reserved.
Connection Track
Connection Slot
1018401 Forehead Support and Pad
1009722 Silicone Forehead Pads, 4 pack
1009723 Ball-and-Socket Swivel Clips (2 pk)
302209 Rotating Connector Package
1002720 Split Washer Kit (5 ea)
1008814 ComfortSelect Retaining Ring
312710 O
Enrichment Attachment
2
1004395 Exhalation Elbow Replacement Kit**
7041
Mask Swivel
332113 Whisper Swivel II
Storage: Temperature -20° C to 60° C
Relative Humidity up to 95%, non-condensing
100 LPM
0.2
Covered by the following U.S. patents: 6,851,425;
Des 483,477; Des, 489, 817. Other patents pending.
Nasal Mask & Premium
Headgear with EZPeel™ tabs
InstructIons for use
Multi-Patient use reusable
Forehead Support
Intended use
The ComfortSelect Nasal Mask and Headgear are intended
to provide an interface for adult patients when used with
Respironics CPAP or bi-level therapy. A separate exhalation
device must be used with this mask.
symbol
The symbol shown below is used on the device.
Attention, consult accompanying documents
Warnings
• A separate exhalation device must be used with this
mask.
• This mask does not include an anti-asphyxia valve that
would allow the patient to breathe if the ventilator
failed.
• This CPAP mask should be used only with CPAP
systems recommended by your physician or
respiratory therapist. A mask should not be used
unless the CPAP system is turned on and operating
properly. The vent hole or holes associated with the
mask should never be blocked. Explanation of the
Warning: CPAP systems are intended to be used
with special masks with connectors which have
vent holes to allow continuous flow of air out of the
mask. When the CPAP machine is turned on and
functioning properly, new air from the CPAP machine
flushes the exhaled air out through the attached
mask exhalation port. However, when the CPAP
machine is not operating, enough fresh air will not
be provided through the mask, and exhaled air may
be rebreathed. Rebreathing of exhaled air for longer
than several minutes can in some circumstances lead
to suffocation. This warning applies to most models
of CPAP systems.
English
Latex free
• At low CPAP pressures, the flow through the
exhalation port may be inadequate to clear all exhaled
gas from the tubing. Some rebreathing may occur.
• At a fixed flow rate of supplemental oxygen flow, the
inhaled oxygen concentration will vary, depending on
the pressure settings, patient breathing pattern, mask
selection and the leak rate. This warning applies to
most CPAP or bi-level systems.
• If oxygen is used with the CPAP or bi-level system,
the oxygen flow must be turned off when the system
is not in operation. Explanation of Warning: When
the CPAP or bi-level system is not in operation and
the oxygen flow is left on, oxygen delivered in the
ventilator tubing may accumulate within the CPAP or
bi-level system enclosure. Oxygen accumulated in the
CPAP or bi-level system will create a risk of fire. This
warning applies to most CPAP or bi-level systems.
• Oxygen accelerates fires. Keep the CPAP or bi-level
system and oxygen container away from heat, open
flames, any oil substance, or other sources of ignition.
Do not smoke in the area near the device or the
oxygen container.
• Some users may experience skin redness. If this
occurs, consult your healthcare professional.
Support Arm
Forehead Pad
Swivel Clip
Standard
Elbow
Important!
Hand wash mask before use.
1033265