6 ENGLISH
OXYGEN/PRESSURE PORT CONNECTOR
If pressure readings and/or additional oxygen are required, an Oxygen/Pressure Port Connector is
available (REF 900HC452). Please contact your healthcare provider.
TECHNICAL SPECIFICATIONS
The F&P Vitera Full Face mask has exhaust holes to expel the air that you exhale from the mask. It is
important that these exhaust holes are not blocked by any object. This controlled leak ensures all exhaled
CO
is expelled from the mask.
2
F&P VITERA EXHAUST FLOW INFORMATION
Pressure (cmH
O)
2
Flow (L/min)
Due to manufacturing variations, the exhaust flow rates may vary from the nominal values shown above.
70
60
50
40
30
20
10
0
4
6
8
10
RESISTANCE TO FLOW
Pressure drop through mask at 50 L/min: 0.24±0.15 cmH
Pressure drop through mask at 100 L/min: 0.47±0.15 cmH
INSPIRATORY AND EXPIRATORY RESISTANCE
Inspiratory resistance of the mask (in combination with the Non-Rebreathing Valve) is 1.05±0.15 cmH
Expiratory resistance of the mask (in combination with the Non-Rebreathing Valve) is 0.80±0.15 cmH
The opening to atmosphere pressure of the Non-Rebreathing Valve is <0.56 cmH
The closing to atmosphere pressure of the Non-Rebreathing Valve is <2.16 cmH
MASK DEAD SPACE
Small = 246 cc
SOUND
The sound power level of mask is mask is 29.8 dBA, with uncertainty 2.5 dBA
The sound pressure level of the mask is 21.8 dBA, with uncertainty 2.5 dBA
ADDITIONAL INFORMATION
The F&P Vitera Full Face mask and accessories are not made with natural rubber latex or DEHP.
The mask should be replaced no later than 12 months after the date of first use.
STORAGE
Store dry mask in clean conditions out of direct sunlight. Storage temperature: -20 to 50°C (-4 to 122°F)
TECHNICAL INFORMATION
4
6
8
10
12
24
28
32
36
40
12
14
16
18
20
Pressure (cmH
O)
2
Medium = 271cc
14
16
18
20
43
47
50
53
22
24
26
28
30
O
2
O
2
Large = 322cc
22
24
26
28
30
55
58
60
62
64
O
2
O
2
O
2
O