Maintenance
To ensure the functionality of the device, periodic visual inspection is recommended for:
•
Secure mounting
•
No mechanical damages or corrosion
•
Worn or otherwise damaged cables
•
The potential equalization terminal must be secured against loosening
•
Correct and clearly marked cable connections
Intrinsic safety "i"
•
Observe the valid regulations for the interconnection of intrinsically safe circuits.
•
The instrument is only suitable for connection to certified, intrinsically safe instruments
•
If the intrinsically safe circuit is lead into dust-explosive areas of zone 20 or 21, please make
sure that the instruments which are connected to these circuits meet the requirements of
EPL Da or EPL Db and are certified respectively
8
Electrostatic charging (ESD)
In case of instrument versions with electrostatically chargeable plastic parts, the danger of electro-
static charging and discharging must be taken into account!
The following parts can charge and discharge:
•
Lacquered housing version or alternative special lacquering
•
Plastic housing, plastic housing parts
•
Metal housing with inspection window
•
Plastic process fittings
•
Plastic-coated process fittings and/or plastic-coated sensors
•
Connection cable for separate versions
•
Type label
•
Isolated metallic labels (measuring point identification plate)
Take note in case of danger of electrostatic charges:
•
Avoid friction on the surfaces
•
Do not dry clean the surfaces
The instruments must be mounted/installed in such a way that the following can be ruled out:
•
electrostatic charges during operation, maintenance and cleaning.
•
process-related electrostatic charges, e.g. by measuring media flowing past
The warning label indicates danger:
VEGAMET 841, 842, 861, 862
WARNING - POTENTIAL ELECTROSTATIC
CHARGING HAZARD - SEE INSTRUCTIONS
IECEx