Replacing The Filter; Storage - Graham Field JOHN BUNN Neb-u-Tyke BELLA the Butterfly Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Replacing the Filter

s NOTICE: Change the filter once every 30 days or when the
filter changes color.
Info: Additional filters can be purchased at
www.grahamfield.com or from your Graham-Field authorized
distributor.
1. Ensure that ON/OFF switch is in
the OFF position. Open the filter
cap located on the bottom of the
nebulizer.
2. Remove the dirty filter with a small,
pointed object. Discard the filter.
3. Replace with a clean filter.
4. Replace the filter cap.
s NOTICE: Do not use cotton or other materials as a filter.
s NOTICE: Do not wash or clean the filter.
s NOTICE: Do not reuse the filter. Do not operate
compressor without a filter. Compressor damage can
result.
WARNING: The disposable nebulizer kit is for single
patient use only.

Storage

Keep the device and accessories dry. Avoid exposure to
direct sunlight.
JB0112-071-INS-LAB-RevB14
filter compartment
(shown without filter)
13

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Jb0112-071

Tabla de contenido