Then rinse all parts thoroughly (for a least 1 minute) under warm
running water (approx. 40 °C without dishwashing liquid) to
remove all remaining dishwashing liquid.
Pour out the water that has collected in the individual parts.
You can remove excess water more quickly by shaking the parts.

Disinfection

Disinfect the dismantled nebuliser and the mouthpiece or the mask
used immediately after cleaning.
Disinfection with boiling water
Place the individual parts of your INQUA
water for at least 15 minutes.
Use a clean pot and fresh drinking water.
Drinking water contains dissolved particles (e.g., salts) that can form
a milky film on plastic parts during the boiling process, depending on
their type and quantity. This film can be prevented by using distilled
water in the boiling operation.
When boiling, make sure that there is enough water in the pot to pre-
vent the nebuliser parts from coming into contact with the base of the
pot.
WARNING: Risk of fire if boiling operation is not
supervised!
As soon as the water has evaporated completely, the plastic
in the pot will start to melt. If the heat source is not cut off, the
plastic will decompose, giving off smoke, and may cause a
fire.
Never leave the boiling procedure unattended!
46
®
nebuliser in boiling