SPARE PARTS/ACCESSORIES
Applied part
Item description
®
INQUA
nebuliser set for INQUA
TECHNICAL DATA
General
Recommended nebuliser gases:
Minimum flow:
Corresponding operating pressure:
Maximum flow:
Corresponding operating pressure:
Minimum fill volume:
Maximum fill volume:
Aerosol characteristics in accordance with EN 13544-1
appendix CC
The output of the nebuliser depends on the medication being used.
Consequently, the following data may not apply for medications that
exist as suspensions or are very viscous. If in doubt, ask the manu-
facturer of the medication concerned for advice.
Determination of particle size
®
Fill the INQUA
attach the INQUA
tion Impactor™) and run for 3 minutes.
nebuliser with 3 ml of a 2.5% test solution and then
®
nebuliser to an impactor (NGI™ = Next Genera-
®
Inhalator
Air
(3.0 l/min)
50 kPa (0.5 bar)
(6.0 l/min)
200 kPa (2.0 bar)
2 ml
8 ml
Item no.
BR021200
51
en