RS232 interface
- Connection
RXD (Input)
TXD (Output)
GND (Ground)
- The baudrate by default is 1200 bps (see Baudrate setting)
- Data format: 10 bits, 0 as start bit, 1 as stop bit, 8 digits (ASCII code)
- No odd and even numbers adjusting
- Data output: by default is continuous mode. The data ouput mode can be changed
into press output, timing output and continuous output (see Data output setting)
- Output data format
1
2
3
Space
type
Space
or *
5. MAINTENANCE AND CLEANING
To get the best results and a higher duration of the balance it is essential to
follow the processes of use.
Note: All the processes of use mentioned below will not have any value unless
you keep a continued and careful maintenance.
•
Please follow the processes of use of this manual.
•
This manual should be available for all users of this equipment.
•
Prevent the balance from sudden movements and knocks, as well as from direct
sunlight or air flows. The balance is a precision instrument, you must handle it
carefully.
•
The balance has to be plugged to an earth connection and the socket should be
handy and ready to unplug the equipment in case of emergency.
•
Never unplug the balance by pulling the wire, do it from the base.
•
Never use the balance in a wedged location as for example a shelf.
•
Never use sharp objects as pens, etc... to press the buttons of the control panel;
only use your fingers.
•
Never place on the pan an object heavier than the maximum capacity of the
balance, the sensor could be damaged.
Balance (9 pins)
2
3
5
4
5
6
7
±
data
data
data
8
9
data or dot
data or dot
28
PC/Printer (9 pins)
3
2
5
10
11
12
13
data
data
data
Unit1
14
15
16
Unit2
end
return