Descargar Imprimir esta página

FujiFilm VISUALSONICS Vevo MD Manual Del Usuario página 13

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

START
RA072 RA075 RA126 RA138 RA139
1 Put on gloves, for protection from cleaning and disinfecting materials. Change
gloves as needed throughout the procedure.
2 Turn off the system by pressing the Power button.
3 Turn off the main power switch located on the bottom of the system near the back,
where the power cord connects.
4 Unplug the power cord from the outlet.
5 Unplug the network cable, if applicable.
6 Disconnect and remove the transducer from the system.
7 Remove the transducer sheath (probe cover), if applicable.
8 Obtain a compatible cleaner. Compatible cleaning materials are listed in the table
Compatible cleaners and disinfectants
Caution:
Make sure the solution does not get into any of the connector ports.
Apply the disinfectant solution to the cloth rather than applying it directly to
surfaces if not using a wipe.
9 Clean the transducer cable and transducer body to remove any visible gel, debris,
or bodily fluids.
♦ Starting from the transducer connector end, clean the transducer cable and
transducer body in a downward motion toward the scanhead. Wipe from clean areas
to the soiled areas to avoid cross-contamination.
Caution:
Do not wipe the transducer connector.
Semi-critical cleaning and disinfection
on page 11.
Cleaning and Disinfecting - English
13

Publicidad

loading