Cannot be immersed
Can be immersed
Connector
Cable
Scanhead
5 Rinse by soaking the area that was immersed, in sterile or clean water, according
to the disinfectant manufacturer's instructions. See the table
RA072 RA0126
Compatible high-level disinfectants for semi-critical contact
on the previous page for a
list of clean water soak times.
♦ For example, Cidex OPA recommends a minimum 1 minute soak repeated 3 times.
This requires either replacing the water three times or having 3 bins with fresh water.
6 Dry with a clean lint-free cloth, or allow to air dry completely.
7 Examine the transducer and cable for damage, such as cracks or splitting where
fluid can enter. If damage is evident, discontinue use of the transducer, and contact
FUJIFILM SonoSite or your local representative.
DONE
Cleaning and Disinfecting - English