Compatible high-level disinfectants for semi-critical contact
Transducer
Disinfectant
UHF22
UHF48
Cidex
UHF70
1 Ensure you have cleaned the transducer as seen in
and transducer, disinfecting of system for semi-critical contact
2 Obtain compatible high-level disinfecting materials. Compatible materials listed in
the table
RA072 RA0126 Compatible high-level disinfectants for semi-critical contact
above.
Caution:
Ensure that you select the proper disinfectant according to your
transducer.
Cidex OPA is not compatible with any FUJIFILM SonoSite UHF transducer.
Use only FUJIFILM SonoSite recommended disinfectants listed in the table
RA0126 Compatible high-level disinfectants for semi-critical contact
damage to the transducer.
3 Follow the information in the table
disinfectants for semi-critical contact
Refer to the disinfectant manufacturer's instructions for additional information.
4 Immerse the transducer into the high-level disinfectant solution. Refer to the table
RA072 RA0126 Compatible high-level disinfectants for semi-critical contact
soak times. Ensure that the transducer connector remains out of the fluid. Ensure that
the scanhead and only 2 in (5 cm) are immersed. Refer to the figure below.
16
Cleaning and Disinfecting - English
Disinfectant
Temperature
soak duration
25°C (77°F)
45 minutes
RA072 RA0126 Compatible high-level
above for solution strengths and soak times.
Clean water
soak time
3 x 1 minute
Part 1 of 2: Cleaning of system
on page 11.
above to avoid
RA072
above for