Transcend 365 miniCPAP Auto Quick Guide
Caring for the Humidifier Reservoir
Follow these instructions to care for the humidifier reservoir.
1. Rinse the reservoir with distilled water weekly.
2. Reattach the humidifier reservoir to the PAP device.
3. Press and hold the Up and Down buttons for three seconds to start Drying
Mode. This mode will last 30 minutes unless you interrupt it by pressing any
button. After the 30-minute drying cycle, the blower will turn off and the
device will automatically enter Standby Mode. Dry mode flushes your CPAP
device, tubing and mask of any residual moisture.
4. Refill the reservoir with water prior to next use.
Replacing the Filter Media
During therapy, ambient air is drawn into
the Transcend 365 miniCPAP Auto through
an Air Inlet Filter. The Filter Media is very
simple to replace. Simply turn the unit
over and look for the door on one end that
pops open easily.
Open door to remove the used filter
media, replace with new filter media and
secure the filter door. The filter media
should be replaced after 6 months of use
or sooner if it is dirty.
Questions? Contact Somnetics customer service 877-621-9626 or
Appendix: Part Numbers
Disposable parts
Accessories
Replacement parts
Flip
open
door
Remove
filter