Temperature Taking Hints
• External factors may influence accuracy, including when an individual has:
– been recently swimming or bathing.
– been lying on the side where the temperature is taken from. Never
take temperature from the side on which the patient was laying.
Use the opposite side.
– been recently exposed to hot or cold air temperatures. Wait at least
20 minutes after the patient was moved to a room with normal
operating conditions.
• Readings from the right side of the patient's head may differ from
readings taken on the left side of the head. Therefore, preferably take
all temperatures on the same side of the head.
Memory Mode
The Vicks Behind Ear Gentle Touch Thermometer stores the last 8
temperature readings. After turning power on, the last temperature is
always briefly displayed.
• To retrieve the last 8 readings, press the "M" Button
when the thermometer has <OK> shown on the Display.
The Display will show the most recent temperature with
the corresponding reading number. Pressing the "M"
Button will show additional readings.
• The Memory Mode is exited either automatically if the "M" Button is no
longer pushed or by pressing the Power Button.
Care and Cleaning
• If the Sensor needs to be cleaned or disinfected,
push and hold the Start Button to move the tip out
of the Washable Ring and gently wipe the Sensor
and the Washable Ring with a cotton swab or piece
of cotton dampened with alcohol. Allow to dry for at
least 5 minutes before measuring temperature.
• Use a soft cloth to clean the thermometer display and exterior. The cloth
can be slightly dampened with water or alcohol.
6
• Do not use abrasive cleaners. Never submerge this thermometer in
water or any other liquid.
• Store thermometer in a dry location at room temperature free from dust
and contamination and away from direct sunlight.
• Do not put it under running tap water or immerse it in the water.
Trouble-Shooting
Problem
Possible Cause
Power can not be turned on Battery is low or not
snapped into position inside
the battery holder.
Product doesn't turn off
Internal malfunction
Patient or thermometer
was moved from a cold
environment.
Thermometer probe was
recently wet cleaned.
Patient or thermometer
was moved from a warm
environment.
Patient was laying on
the same side where
measurement was taken.
Internal malfunction
Thermometer was moved during
temperature taking process
Thermometer was not in the
correct position during the
temperature taking process
Short measurement time
Start button was pressed
too slow
Solution
Make sure battery is tightly
snapped into battery holder. If the
battery is dead, replace battery.
Remove and replace battery
Wait at least 20 minutes before
retaking temperature.
Wait at least 5 minutes after cleaning
before retaking temperature.
Wait at least 20 minutes before
retaking temperature.
Measure temperature from the
opposite side.
Restart thermometer. If this
doesn't fix the problem,
contact Kaz.
Hold thermometer still and
retake measurement
Make sure orange ring of
thermometer is behind ear lobe
and retake measurement
Make sure that the Start Button is
kept pressed until the beep sounds
Retake the temperature and
ensure the start button is
pressed down quickly
7