5. Equipment calibration checking
By using a standard resistance (Rx), it is possible to test the equipment
calibration. The resistance must be connected to the EM-4058 as shown
below:
After connecting the resistance, follow these procedures:
Turn the EM-4058 on (
·
Select the resistivity function (
·
In the distance / standard resistance selection function, choose Rx
·
(nominal standard resistance).
Press the
·
After 5 seconds, a stable value close to the standard resistance's
·
nominal value should appear on the display.
If this is not the case, the equipment accuracy is out of its
·
specification and must be adjusted by a qualified laboratory.
).
key.