Guidance and manufacturer's declaration
Guidance and manufacturer's declaration - Electromagnetic emissions
The device Kinetec Spectra Essential™ is intended for use in the electromagnetic environment specified below. The customer or the user of
the device Kinetec Spectra Essential™ should ensure that it is used in such an environment.
Emissions test
Radio frequency emissions CISPR 11
Radio frequency emissions - CISPR 11
Harmonic emissions - IEC 61000-3-2
Voltage fluctuations / Flicker emissions - IEC
61000-3-3
Guidance and manufacturer's declaration - Electromagnetic immunity
The device Kinetec Spectra Essential™ is intended for use in the electromagnetic environment specified below. The customer or the user of
the device Kinetec Spectra Essential™ should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient /
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage interruptions
IEC 61000-4-11
Voltage dips and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency (50/60
Hz) magnetic field - EC
61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test level.
T
Kinetec Spectra Essential™
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
Test level
±2 kV, ±4 kV, ±6 kV
contact
±2 kV, ±4 kV, ±8 kV
air
±2 kV
for power supply lines
±0,5 kV, ±1 kV
between lines
±0.5 kV, ±1 kV, ±2 kV
between line and earth
< 5% U
T
(>95% dip in U
)
T
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
3A/m
Electromagnetic environment – guidance
The device Kinetec Spectra Essential™ uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The device Kinetec Spectra Essential™ is suitable for use in all establishments
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic purposes.
Compliance level
Electromagnetic environment – guidance
±2 kV, ±4 kV, ±6 kV
contact
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
±2 kV, ±4 kV, ±8 kV
should be at least 30%.
air
±2 kV
Mains power quality should be that of a typical
for power supply lines
commercial or hospital environment.
±0,5 kV, ±1 kV
between lines
Mains power quality should be that of a typical
±0.5 kV, ±1 kV, ±2 kV
commercial or hospital environment.
between line and
earth
< 5% U
When the interruption occurs, the device Kinetec
T
(>95% dip in U
)
Spectra Essential™ is reset. After turning on, push START
T
for 5 seconds
to begin the session.
< 5%
U
T
(>95% dip in U
)
T
Mains power quality should be that of a typical
for 0.5 cycle
commercial or hospital environment. If the user of the
device Kinetec Spectra Essential™ requires continued
40% U
T
operation during power supply interruptions, we
(60% dip in U
)
T
recommend powering the device Kinetec Spectra
for 5 cycles
Essential™ using an uninterruptible power supply or a
battery.
70% U
T
(30% dip in U
)
T
for 25 cycles
Power frequency magnetic fields should be at levels
3A/m
characteristic of a typical location in a typical
commercial or hospital environment.
EN