General Setup (Applies For All Parameters) - Mettler Toledo M400 Guia De Inicio Rapido

Ocultar thumbs Ver también para M400:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 99
5

General Setup (applies for all parameters)

21.7
%sat
25.0
°C
MENU
Quick Setup
Channel Selection
6.0
A
pH
25.0
A
°C
Channel Select = ISM
Parameter = Auto
Analog out puts
7.0
pH
25.0
°C
Aout1 min= 0.000 pH
Aout1 max= 14.00 pH
h
10
While in Measurement mode press the [MENU] key to bring up the Menu selection.
S elect QUICk SETUP and press the [ENTER] key.
Display Convention:
1st line on display
2nd line on display
Select the units of measurement for a and b. Only lines a and b can be configured
in QUICk SETUP. Go to the Configuration Menu to configure lines c and d.
Please select the type of Sensor:
Analog: For conventional analog sensors (will be displayed on channel "A").
ISM:
For ISM sensors (will be displayed on channel "B").
Please select the parameter requirement:
The choice of parameter depends on the level of transmitter. If an ISM sensor is se-
lected, the setting "Auto" means, all possible ISM sensors will be recognized and
accepted. If a special parameter is chosen, only this parameter will be recognized
and accepted on the transmitter.
By using analogue 2-e or 4-e conductivity sensors,
see "Conductivity Quick Setup" below for intermediate steps.
By selecting YES the linear 4–20 mA analog output Aout1 will be set up for measure-
ment a when [ENTER] is pressed. Selecting NO means that no analog output is set
up.
Aout1 min, Aout1 max are the minimum and maximum measurement values for the
u
4 and 20 mA values respectively.
Note for multi channel devices: The user can configure output A out 3 to mesure-
ment c by going back to the previous menu and selecting c.
(PATH: Menu / Quick Setup)
a
3rd line on display
b
4th line on display
c
d

Publicidad

Tabla de contenido
loading

Tabla de contenido