Calibration
Calibration to Percent Saturation (SAT), in Water
Display
10
Action
Press cal key, enter code 1100.
Select with key,
edit number with key,
proceed with enter.
Immerse sensor in cal medium
Start with enter.
Enter relative humidity
Select using,
enter number using .
Confirm entry with enter.
Enter calibration pressure
Select using,
enter number using .
Confirm entry with enter.
Automatic drift check
Display of sensor current (related
to 25 °C and 1013 mbars nor-
mal pressure) and measuring
temperature.
The drift check might take some
time.
Remark
SAT or Conc calibra-
tion is selected during
configuration.
If an invalid code is
entered, the device
returns to measuring
mode.
Device is in the Hold
mode.
Default for relative
humidity in aqueous
media:
rH = 100 %
(in air approx. 50 %)
Default for calibration
pressure is the process
pressure configured
Drift check can be
stopped after > 10 sec
by pressing cal (accu-
racy reduced).