Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Please contact the place of purchase or call the Customer
Service for assistance.
NOTE:
• The control solution range printed on the test strip vial is for control solution
use only. It is not a recommended range for human blood glucose level.
• See the Maintenance section for important information about your
control solutions.
4. Read and compare the result
After counting down to 0, the control
solution test result will appear on the
display. Compare this result with the
range printed on the test strip vial and it
should fall within this range. If not, please
read the instructions again and repeat
109~165
the control solution test.
(136 mg/dL = 7.6 mmol/L; 109-165 mg/
dL = 6.1-9.2 mmol/L)
21