EN
5 Specifications
Mains connection
Supply voltage
Consumed power – max.
Fuses
NF leakage current
In accordance with
Leakage currents against Earth
Leakage currents against Housing
Patients Leakage currents
Patients – auxiliary stray currents
HF leakage currents – measured in a system with a electrosurgical device
In accordance with
Neutral electrode against Earth
Aktive electrode against Earth
Mode with Argon
Argon Minimum purity
Argon gas sources
Control of the output on the exit
cylinder
Exchange empty
Safety measures
Type of the construction according to
Protection class
Type
Earthing terminal
Switching-on the neutral electrode
•
Monitoring for wrong dosing of the output and switching-off the hf – generator.
•
Monitoring of the fault conditions and switching-off the delivered output and hf – energy
32
cylinder
with full
100-260 Vac, 50-60 Hz
18 VA
2 St. T 1,0 A (inerte)
EN 60601-1; IEC 60601
< 500A
< 1000A
< 10A
< 100A
EN 60601-2; EN 60601
< 100 mA
< 100mA
99.998 %
•
1 or 2 pieces gas
maximum pressure 200bar and
recommended filling volume 25 l
•
Gas plant with highest inlet pressure
2,5bar
0,3 – 6,0 l/min with Step 0,1 l/min regardless for
both operating modes Cut and Coagulation
At output on entrance of the System <9 l/min
EN 60601; IEC 60601
I
CF
yes
Floating Output
HBS Argon
cylinders
with