ELECTRICAL CONNECTION

The TCW2043SDD is intended to be connected to a double b medical systems Solar generator
system.
Each unit (refrigerator and freezer) is designed to work with a minimum solar radiation reference
period of 3.5 kWh/m2/day according to WhO-specification.
Each generator must provide a voltage range of 12V to 20V DC with a minimum power of 320W at
STC.
before connecting the TCW2043SDD and switching it on for the first time, allow it to stand for 30
minutes.
CONNECTING TO ThE b MEDICAL SySTEMS SOLAR GENERATOR SySTEM
For installation of the solar power generator refer to the Installation instructions.
Connect the TCW2043SDD to the generators with the provided plug system.
6