OPERATION
TEMPERATURE-DISPLAy
control of the refrigerator
•
The TCW2043SDD and its control panel is composed of 2 separate units : the refrigerator part on the
left side, and the freezer part on the right side of the control panel.
•
In normal use, the display shows the inside temperature of the appliance. This temperature is
displayed in °C with one decimal place.
•
During times without solar energy supply or in case of power failure the display is switched off. The
inside temperature can be checked by pressing the "TMP" button.
TEMPERATURE-SETPOINTS
•
The temperature setpoints of the refrigerator- and the freezer unit are preset at the factory at
following values: +5°C for the refrigerator, -25°C for the freezer.
•
These temperature setpoints cannot be changed by the user.
•
Press the SET-key to view the actual temperature setpoint.
STORAGE OF VACCINES IN ThE REFRIGERATOR
IMPORTANT :
When using the TCW2043SDD refrigerator, it is of utmost importance that the icelining
waterpacks are in place. Make sure that all waterpacks are in place before starting the
TCW2043SDD in refrigerator mode. Never remove the waterpacks from the unit.
Special care should be taken to preserve the vaccines that get damaged by freezing. Never store
them touching the side surfaces in the cabinet of a TCW2043SDD. It is advisable to preserve the
vaccines in the provided baskets.
For storing vaccines :
•
keep the packets containing the vaccines in
neat rows.
•
Different vaccines should be kept separately to
facilitate easy identification.
•
keep about 2 cm space between rows for
circulation of air.
•
Do not load vaccines higher than the indicated
maximum level.
REFRIGERATOR - REFRIGERATEUR - REFRIGERADOR
SET
TMP
RST
FREEZER - CONGELATEUR - CONGELADOR
SET
TMP
control of the freezer
7
RST
maximum
vaccine load