Lovibond Tintometer MD 100 Manual De Instrucciones página 15

Ocultar thumbs Ver también para Tintometer MD 100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 147
Ammonia T / M60
Ammonia T
0.02 - 1 mg/l N
Indophenole Blue
Material
EN
Required material (partly optional):
Reagents
Ammonia No. 1
Ammonia No. 1
Ammonia No. 2
Ammonia No. 2
Set Ammonia No. 1/No. 2 100 Pc.
Set Ammonia No. 1/No. 2 250 Pc.
Ammonia Conditioning Powder
Preparation
1. Sea water samples:
Ammonia conditioning reagent is required when testing sea water or brackish water
samples to prevent precipitation (settlement) of salts.
Fill the test tube with the sample to the 10 ml mark and add one level spoonful of
Aluminium Conditioning Powder. Close the vials with the caps and swirl until the
powder has dissolved. Then proceed as described.
Notes
1. The AMMONIA No. 1 tablet will only dissolve completely after the AMMONIA No. 2
Tablet has been added.
2. The temperature of the sample is important for full colour development. At
temperatures of below 20 °C the reaction period is 15 minutes.
16
Packaging Unit
Tablet / 100
Tablet / 250
Tablet / 100
Tablet / 250
100 each
#
250 each
#
Powder / 15 g
EN   Method Reference Book   1.0
M60
A
Part Number
512580BT
512581BT
512590BT
512591BT
517611BT
517612BT
460170