Descargar Imprimir esta página

Lovibond Tintometer MD 100 Manual De Instrucciones página 7

Ocultar thumbs Ver también para Tintometer MD 100:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 35
EN
Dissolve tablet(s) by
inverting.
12
11
10
9
8
7
6
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/l free Copper appears on the display.
Implementation of the provision Copper, total with tablet
Select the method on the device
In addition, choose the test: total
Fill 24 mm vial with 10 ml
sample.
8
Copper T / M150
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
2 min
1
2
3
4
5
10 ml
Close vial(s).
Sample
Press the TEST (XD:
START)button.
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
EN   Method Reference Book   1.0
Test
Sample

Publicidad

loading

Este manual también es adecuado para:

Tintometer md 110Tintometer md 200