2. Use
When the heater is ready for use, up to four HEROfill
appropriate slots. Place the HEROfill
Obturator to be used first in slot #1, the second in #2 etc. - generally you should begin with the
smallest size HEROfill
Obturator.
®
1. Activate the timer by pressing the ON icon (B4).
2. Indicator B lights up red during the heating cycle.
3.
When ready, the indicator changes to green and you will hear a beep (B4).
Let the HEROfill
Obturator stay in the heater until you are ready to insert it (max. 15 min).
®
4. It is possible to turn off the HEROfill
seconds (B4). The HEROfill
button for 2 seconds to reactivate the heater.
Tip: You do not have to turn off
be left on standby all day.
Indicators - light and sound
3. Cleaning
When the heater has cooled completely, gutta percha residue may easily be removed. The top of the
heater may be removed for cleaning by rotating the top counter-clockwise 1 until the top is loose
and can be removed 2 (B5).
The top may be washed in a dishwasher and/or autoclaved. Gutta percha residue may be removed
from the heat chamber by the use of a blunt plastic instrument.
The main part of the heater may be wiped with a slightly damp cloth, using standard cleaning
solutions and surface disinfection products.
Sterilization guide for removable top part
WARNINGS
Limitations on reprocessing
Instructions
Point of use
Preparation for decontamination
Cleaning: Automated
Cleaning: Manual
Obturators in slots marked 1 to 4 (B3). Tip: Place the HEROfill
®
Oven by holding down the ON button for 2
®
Oven beeps three times to indicate it is switched off. Hold down the ON
®
the HEROfill
Oven after finished treatment; the HEROfill
®
Obturators at a time may be placed in the
®
DISCHARGE DAMAGED PRODUCTS
250 cycles
Remove excess soil with disposable cloth/paper
tissue.
No specific requirements
Not recommended
Equipment: Detergent, brush, distilled or
demineralised water.
®
Oven can
®
7