7 Fault conditions

7.1

On/off lamp does not light and the paraffin wax remains cold

Check that the mains cable is correctly connected to the unit and to the wall socket
Check that the wall socket is live (for instance with another electrical unit)
Check that the mains cable is not defective (use a spare cable)
Check the mains fuses
7.2

Replacing mains fuses

Switch off the apparatus, remove the mains plug from the wall socket and disconnect the plug at the unit
input. Unscrew the fuse holder with a coin and replace the fuse with one of the same type (6,3 AF type
G). Screw the fuse holder in and reconnect the mains cable.
7.3

Warning lamp fault conditions [7] lights

This indicates that the unit is overheated, probably as a result of a technical defect. Switch off the unit and
remove the mains plug from the wall socket. Contact your local supplier. The Fango Heater can be
cleaned with a damp cloth and a (non abrasive) liquid household cleaner.
In order to obtain spare parts, you please contact your supplier.

8 Finally

Disposal
Your Fango Heater contains materials which can be recycled and/or are noxious for the environment.
At the end of the duration of life, specialised companies can take apart the Fango Heater and sort out the
noxious materials and materials for recycling. By doing so you contribute to a better environment.
Please ensure that you are well informed of the local rules and regulations regarding to the removal of
equipment and accessories.
We hope that you will use the Fango Heater for a long time with much satisfaction and we take the liberty
of drawing your attention to the many other high quality products in our programme. If you should have
any questions or wish to receive our catalogue, please contact our representative or visit us at
www.enraf-nonius.com.

9 Technical data

Mains voltage
Current consumption
Range of thermostat
Sterilization temperature
Safety device
Capacity
Dimensions
Safety Class (IEC 60601-1)
Technical modifications reserved.
: available for 220 – 240 V, 50 – 60 Hz
: 1200 W
: 0 – 100º C
: 135º C
: automatic switch-off when overheated (160º C)
: 15 l
: 37 x 37 x 41,5 cm
: I typ BF