2. Guide to using the Comp Air Elite

Preparing the Nebuliser for Use (Section 3)
Note: Clean and disinfect the nebuliser kit and applicable inhalation
attachments before using the device for the first time, or if it has not
been used for more than one week.
Taking Medication (Section 4)
Notes:
• Do not tilt the nebuliser kit so that it is at an
angle of greater than 45
• Do not crease or fold the air tube.
Cleaning Parts After Use (Section 5)
Note: Failure to clean the nebulising parts
after use may lead to infection.
How the nebuliser kit works
The compressed air from the
compressor is forced through the
nozzle, as it does so, it is mixed with
the medication in the vaporiser head
and forced upwards. The medication
is then turned into a fine aerosol and
directed upwards by the baffle,
where it is broken down into many
smaller aerosol particles.
10
°
.
Aerosol
Nozzle
Medication
Compressed Air
Baffle
Aerosol
Vaporiser
Head
Medication