Troubleshooting - Omron Comp Air Elite Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 146
NE-C30_main.book Page 24 Friday, October 3, 2008 11:46 AM

8. Troubleshooting

Check the following if your unit should fail during operation. You can also refer to
the pages of this manual for complete instructions.
Symptom
Nothing happens when the
power switch is pressed.
(The power indicator does
not light.)
No nebulisation or low
nebulisation rate, when the
power is on. (The green
power indicator is lit.)
The device is abnormally
loud.
24
Cause
Is the AC adapter connected
correctly to the an electrical
outlet and the compressor?
Is the optional rechargeable
battery connected correctly?
Is the optional rechargeable
battery charged sufficiently?
Is the optional DC auto
adapter connected
correctly?
Is there medication in the
medication tank?
Is there too much/little
medication in the medication
tank?
Is the vaporiser head
missing and not assembled
correctly?
Is the nebuliser kit
assembled correctly?
Is the nozzle blocked?
Is the nebuliser kit tilted at a
sharp angle?
Is the air tube connected
correctly?
Is the air tube folded or
damaged?
Is the air tube blocked?
Is the air filter dirty?
Is the air filter cover attached
correctly?
Remedy
Check that the plug is
inserted in an electrical
outlet. Unplug then reinsert
the plug if necessary.
Make sure that the battery is
correctly attached.
Charge the battery.
Make sure that the adapter
is correctly attached.
Fill the medication tank.
Add the correct amount of
medication to the medication
tank.
Attach the vaporiser head
correctly.
Assemble the nebuliser kit
correctly.
Make sure that the nozzle is
free of blockages.
Make sure that the nebuliser
kit is not tilted at an angle of
more than 45 degrees.
Make sure that the air tube is
correctly connected to the
compressor and nebuliser.
Make sure that the air tube
does not contain kinks.
Make sure that the air tube is
free of blockages.
Replace the air filter with a
new one.
Attach the air filter cover
correctly.

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Ne-c30-e

Tabla de contenido