ELECTROMAGNETIC COMPATIBILITY INFORMATION

Table 1 - Emission
Phenomenon
Compliance
CISPR 11
RF emissions
Group 1, Class B
Harmonic
IEC 61000-3-2
distortion
Class A
Voltage fluctuations
IEC 61000-3-3
and flicker
Compliance
Table 2 - Enclosure Port
Phenomenon
Basic EMC standard
Electrostatic
IEC 61000-4-2
Discharge
Radiated RF EM field
IEC 61000-4-3
Proximity fields from
RF wireless communi-
IEC 61000-4-3
cations equipment
88
Visiomed - KD-795 - Manuel d'utilisation - 022019
Rated power
frequency magnetic
Electromagnetic environment
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency
Home healthcare environment
Immunity test levels
Home Healthcare Environment
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Refer to table 3
IEC 61000-4-8
fields
Band (MHz)
(MHz)
385
380-390
450
430-470
710
704-787
745
780
810
870
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
Visiomed - KD-795 - Manuel d'utilisation - 022019
30A/m
50Hz or 60Hz
Immunity test levels
Professional healthcare
facility environment
Pulse modulation 18Hz, 27V/m
FM, ±5kHz deviation, 1kHz sine,
28V/m
Pulse modulation 217Hz, 9V/m
Pulse modulation 18Hz, 28V/m
Pulse modulation 217Hz, 28V/m
Pulse modulation 217Hz, 28V/m
89