CareFusion MicroCo Manual De Funcionamiento página 32

Tabla de contenido

Publicidad

WARNING:
The MicroCO should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use with other equipment is necessary,
the MicroCO and the other equipment should be observed /monitored, to verify
normal operation in the configuration in which it will be used.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuatuins
/ flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroCO is intended for use in the electromagnetic environment specified
below. The customer or the user of the MicroCO should assure that it is used in
such an environment.
Immunity
Test
Electrostatic
discharge
(ESD)
IEC61000-4-2
Electrical fast
transient /
burst
Compliance
Group 1
Group B
Not Applicable
(battery
powered)
Not Applicable
(battery
powered)
IEC 60601 test
level
± 6 kV contact
± 8 kV air
± 2 kV for
power supply
lines
Electromagnetic Enviroment -
Guidance
The MicroCO uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
The MicroCO is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes
Compliance
Level
± 6 kV
contact
± 8 kV air
Not Applicable
(Battery
Powered and
30
Electromagnetic
Enviroment -
Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Mains power quality
should be that of a
typical commercial or

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido