Electromagnetic Compatibility (Emc) To En60601-1-2:2007 - CareFusion MicroCo Manual De Funcionamiento

Tabla de contenido

Publicidad

Electromagnetic Compatibility (EMC) to
EN60601-1-2:2007
WARNING: use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause unexpected or
adverse operation
The MicroCO has been tested to EN60601-1-2:2007, regarding its ability to
operate in an environment containing other electrical/electronic equipment
(including other medical devices).
The purpose of this testing is to ensure that the MicroCO is not likely to
adversely affect the normal operation of other such equipment and that other
such equipment is not likely to adversely affect the normal operation of the
MicroCO.
Despite the testing of the MicroCO that has been undertaken, normal operation
of the MicroCO can be affected by other electrical/electronic equipment and
portable and mobile RF communications equipment.
As the MicroCO is medical equipment, special precautions are needed regarding
EMC (electromagnetic compatibility).
It is important that the MicroCO is configured and installed/put into service,
in accordance with the instructions/guidance provided herein and is used only
in the configuration as supplied.
Changes or modifications to the MicroCO may result in increased emissions or
decreased immunity of the MicroCO in relation to EMC performance.
The MicroCO should be used only with the PC serial cable provided by
Carefusion (Cat. No. 36-CAB1000, Description: Interface cable for COBRA
software). This cable should not be extended by the user. This cable should not
be used with devices other than the MicroCO. If the cable is extended by the
user, this may result in an increased level of emissions or decreased level of
immunity, in relation to the MicroCO's EMC. Use of the cable with devices other
than the MicroCO, may result in an increased level of emissions or decreased
level of immunity, in relation to the other devices' EMC.
The MicroCO meter has an essential performance – the product should continue
to operate as intended, to an accuracy of +/- 5% of reading or 1ppm
(whichever is greater).
63

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido