5 Operation
20
5
Operation
NOTICE! Use only burs and diamond polishers that are sharp and
Use only undamaged burs
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
CAUTION! A loose or partially removed bur can detach itself from the
Drill Fixation _ Turbine
head or break off. This may cause injury! Therefore, use the turbine only
if the bur is inserted at least 10 mm and is clamped securely in place.
CAUTION! Insufficient cooling leads to overheating of the
Overheating at the preparation site
preparation site and damage to the tooth substance. Ensure that the
water content is > 50ml/min.
CAUTION! Do not pull the patient's cheek back with the handpiece
Pulling back the patient's cheek _ Turbine
while the turbine is rotating. This would actuate the pushbutton, thus
creating a risk of burning the patient's oral mucosa.
IMPORTANT: For application details and operating data of the different
Observing manufacturer documentation
preparation tools, please refer to the information provided by the
manufacturer.