Microlife NEB200 Manual Del Usuario página 23

Ocultar thumbs Ver también para NEB200:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment
Immunity Test IEC 60601
Test level
Conduced RF
150KHz to 80MHz:
IEC61000-4-6
3Vrms
6Vrms (in ISM
and amateur radio
bands)
80% Am at 1kHz
Compliance
Electromagnetic
level
environment - guidance
150KHz to
Portable and mobile RF communications
80MHz:
equipment should be used no closer to any
3Vrms
part of the 240391, including cables, than the
6Vrms (in ISM
recommended separation distance calculated
and amateur
from the equation appropriate for the frequency
radio bands)
of the transmitter. Recommended separation
distances:
80% Am at
1kHz
d=1.2√P;
d=2√P
21

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido